Protocol Parameters

Table 7 lists the parameters for temperature and gradient steps with the limits of the
parameters.

Table 7. Protocol steps and parameters of the C1000 Touch thermal cycler.

Step Name

Parameters and Ranges

Description

Temperature

Temperature in °C: The target
temperature between 0.0 and
100.0°C in tenths of a degree

Instructs the thermal cycler to ramp to
the target temperature and hold that
temperature for the specified amount
of time

Hold time: The hold time
between 1 sec and 18 hr in the
format of hr:min:sec. To enter an
infinite hold, touch the

∝ (infinite)

button

Gradient range

Lower: The lower temperature
in the gradient. Enter a number
between 30.0 and 99.0°C in
tenths of a degree

Upper: The upper temperature
in the gradient. The maximum
temperature is 100°C. Enter a
temperature within 24.0°C of the
lower temperature

Instructs the thermal cycler to ramp to
the target temperature gradient across
the block and hold that temperature
gradient for the specified amount of
time

Time: The hold time between 1
sec and 18 hr in the format of
hr:min:sec. To enter an infinite
hold, touch the

∝ (infinite) button

Increment

A temperature from –10.0 to
10.0°C per cycle in tenths of a
degree

Applies only to a temperature step.
Instructs the thermal cycler to
increment (change) the target
temperature of a step with each cycle,
where a positive number increases the
temperature and a negative number
decreases the temperature

Ramp rate

A number from 0.1 to 5°C per
sec

Applies only to a temperature step.
Instructs the thermal cycler to ramp to
the target temperature at the specified
ramp rate in that step

Extend

A time from –60 to 60 sec per
cycle

Applies to both temperature and
gradient steps. Instructs the thermal
cycler to extend the hold time with
each cycle. A positive number
increases the hold time and a negative
number decreases the hold time

Beep

(No parameters)

Applies only to a temperature step.
Instructs the thermal cycler to beep to
signal that the thermal cycler has
reached the target temperature for that
step