 Flow Total Index Enter the reading from the meter at installation
so the total in PrimeLog will be consistent with the meter.
 Low Flows Select this checkbox if the meter will send less than
one pulse per minute.
 Logging Functions Modify the logging function selection to
setup the appropriate function. Average & Total data can be logged 
independently. 
 Show Events Select this checkbox to program the PrimeLog to
record/provide notification based on internal (threshold) events or to 
trigger an output on the logger based on control output events. 
Rain Gauge  This involves logging only when a specific amount of 
rainfall accumulates in a rain gauge tipping bucket. The user must 
designate the amount of rainfall required to cause the tipping 
mechanism in the bucket to tip. 
Pump This involves logging only when a user-defined high or low 
indicates a pump is on or off. The user also must designate the flow rate 
at which the pump turns on. 
Overflow This involves logging only when a user-defined high or low 
indicates an overflow state. 
Analog Screen Description
Select the type of sensor (Transducer/Voltage/Current).
Transducer Enter the PCal for the transducer. Select the Autozero 
on Program checkbox to remove the atmospheric offset when 
programming the logger. 
Voltage Enter the minimum scale and maximum scale values in user 
units and the offset in volts that the minimum scale represents. 
Current Enter the minimum scale and full scale values in user units 
and the offset in mA that the minimum scale represents. 
Note: When using analog inputs on a PrimeLog, use the correct cable 
to connect to the logger source. Current and voltage cables cannot be 
interchanged. Improper use could cause internal damage to the logger. 
It also is important to use a voltage cable of the appropriate rating. Do 
not apply a 10-volt source using a 2-volt or 5-volt cable. This will 
overload the logger. 
The user can select an application for all analog sensors. This will determine 
what is logged by the logger. When using a depth transducer to measure open 
channel flow, both the depth and the derived flow will be saved during 
readback. 