8 metrohm solvent delivery unit, 1 metrohm sdu features – Metrohm IC Net 2.1 User Manual

Page 88

background image

6 Devices

IC Net 2.1

80

Advanced

The

Advanced

tab of the

709 IC Pump

window is only available for a

connected system.

Correction factor

Flow correction factor to adjust the dis-
played flow to the actual flow measured
with the aid of a measuring cylinder.
Entry range:

0.90 ... 1.10

Read

current

Correction factor

value.

Send

current

Correction factor

value to the

709 IC Pump. This parameter is not stored
in the system file (

*.smt

) as long as the file is

not saved.







6.8

Metrohm solvent delivery unit

6.8.1

Metrohm SDU features

A Metrohm SDU (Solvent Delivery Unit) system can be used for
high-pressure gradient mixture of one to four eluents. For each
eluent, a 709 IC Pump is used. Each 709 IC Pump must be con-
nected to the 762 IC Interface using a 6.2134.080 or 6.2134.090
cable. Examples for interconnections with the 709 IC Pump can
be found in the 762 Instructions for Use.
The outlet capillaries of the 709 IC Pumps must be connected to
a mixing chamber from where one inlet capillary leads to the in-
jection valve of the 733 IC Separation Center.
For detailed information about the 709 IC Pump see 709 Instruc-
tions for Use
.