6 Devices

IC Net 2.1

66

Alarm stops

The

Alarm stops

tab of the hardware settings window defines the

events for which the 761 Compact IC is stopped immediately. At
an alarm stop, high-pressure pump and peristaltic pump are
stopped immediately, the running determination and the active
sample queue are also stopped.

Event

Events for alarm stop:

Pump stop

Pump stopped because pressure limits are
exceeded.

Hardware error

Hardware error detected at the 761 Com-
pact IC (high-pressure pump, injection
valve, or suppressor not working correctly).

Leak

Leak detector has detected solvent in the
instruments interior. This information is also
stored in the instrument itself, so that it is
stopped automatically even without con-
nection to the PC.

Power on

Temporary power failure at the 761 Com-
pact IC hardware.