1.4 Installation
IC Net 2.1
5
1.4.3 Software
update
For software update from IC Net 2.0 to IC Net 2.1, proceed as fol-
lows:
1. Switch on PC and start operating system.
2. If you want to save chromatograms, methods and system files
created by IC Net 2.0, backup the folders
Data
,
Methods
, and
Systems
in another directory on the hard disk.
Systems of IC Net 2.0 containing 761 Compact IC icons can-
not be used in IC Net 2.1 any more. Delete these systems be-
fore making the backup of the
Systems
folder.
3. If you want to save data created by Autodatabase 1.0, backup
the folders
Filters
,
Report templates
, and
User database
in an-
other directory on the hard disk.
4. Deinstall the IC Net 2.0 software by selecting
/
Settings
/ Control panel
, double-clicking the
Software
icon, selecting
Metrohm IC Net 2.0
in the list and clicking on
. All
program files and icons should be removed.
5. Delete the folder
IC Net 2.0
in the program folder.
6. Install the IC Net 2.1 software (see section 1.4.2).
7. Copy the IC Net 2.0 backup files of the folders
Data
,
Methods
,
and
Systems
into the folders
..\IC Net 2.1\IC Net\Data
,
..\IC Net
2.1\IC Net\Methods
, and
..\IC Net 2.1\IC Net\Systems
.
8. Copy the Autodatabase backup files of the folders
Filters
,
Re-
port templates
, and
User database
into the folders
..\IC Net
2.1\Autodatabase\Filters
,
..\IC Net 2.1\Autodatabase\Report tem-
plates
, and
..\IC Net 2.1\Autodatabase\User database
.