beautypg.com

3 software update – Metrohm IC Net 2.1 User Manual

Page 13

background image

1.4 Installation

IC Net 2.1

5

1.4.3 Software

update

For software update from IC Net 2.0 to IC Net 2.1, proceed as fol-
lows:
1. Switch on PC and start operating system.
2. If you want to save chromatograms, methods and system files

created by IC Net 2.0, backup the folders

Data

,

Methods

, and

Systems

in another directory on the hard disk.

Systems of IC Net 2.0 containing 761 Compact IC icons can-
not be used in IC Net 2.1 any more. Delete these systems be-
fore making the backup of the

Systems

folder.


3. If you want to save data created by Autodatabase 1.0, backup

the folders

Filters

,

Report templates

, and

User database

in an-

other directory on the hard disk.

4. Deinstall the IC Net 2.0 software by selecting

/

Settings

/ Control panel

, double-clicking the

Software

icon, selecting

Metrohm IC Net 2.0

in the list and clicking on

. All

program files and icons should be removed.

5. Delete the folder

IC Net 2.0

in the program folder.

6. Install the IC Net 2.1 software (see section 1.4.2).
7. Copy the IC Net 2.0 backup files of the folders

Data

,

Methods

,

and

Systems

into the folders

..\IC Net 2.1\IC Net\Data

,

..\IC Net

2.1\IC Net\Methods

, and

..\IC Net 2.1\IC Net\Systems

.

8. Copy the Autodatabase backup files of the folders

Filters

,

Re-

port templates

, and

User database

into the folders

..\IC Net

2.1\Autodatabase\Filters

,

..\IC Net 2.1\Autodatabase\Report tem-

plates

, and

..\IC Net 2.1\Autodatabase\User database

.