7 Methods

IC Net 2.1

174

type

The types of components are designated
by a one-letter code:

R

Reference component (used for peak
identification).

S

Standard

component.

C

Calibration standard in the case it is differ-
ent from the quantification standard.

?

Some points of corresponding peak are out
of ADC or detector range – suspicious re-
sult.

!

Component concentration is outside of
minimum (

min C

) and maximum limits

(

max C

), set in the component table.

p

Special component (it has one of the spe-
cial flags checked on in the component
window).

N

The component response is outside of cali-
brated region.
In the case of peaks additional letters pre-
cede the component type information, indi-
cating, how peaks are separated from other
peaks, for example:

BD_

Peak that starts on the baseline (

B

) and

ends on the drop line (

D

) that separates it

from another adjacent peak.

BBR

Special case of a rider (

R

) peak that is tan-

gentially separated from the main one. A
main peak will have a third letter

H

(horse)

in this case.
The complete component type may look
like:

BBD : !R

.

group

Number of the group for the component.

spectral ratio

Ratio of responses on different channels for
multi-channel chromatograms.

name

Name of the component.

file name

File name of chromatogram. This column is
very useful for the purpose of processing of
exported data.

ident

Title (identifier) of the chromatogram. This
column is useful for the processing of ex-
ported data.