6.5 817 Bioscan
IC Net 2.1
73
Method data
Cell potential
Definition of start and end potential for a
user-defined potential sweep, which is ap-
plied to the amperometric detector cell.
E1
Start potential for sweep.
Range:
-2 ... +2 V
.
E2
End potential for sweep.
Range:
-2 ... +2 V
.
Scan cycle
Definition of scan cycle:
full
double
scan
E1 ... E2 ... E1
half
single
scan
E1 ... E2
Selection:
full, half
Scan rate
Definition of scan rate in mV/s.
Selection:
1, 2, 5, 10, 20, 50 mV/s
Oven
Switching on/off column oven and setting
operating temperature.
Range:
15 ... 60 °C
Range
Selection of measuring range:
pA
10 ... 5000 pA
.
nA
0.1 ... 50 nA
.
µA
0.01 ... 5 µA
.
Full
scale
The full scale range (operating range) sets
the desired sensitivity for the analog output.
The possible values of the full scale range
depend on the preset measuring
Range
.
Offset
Offset of the zero point for the analog out-
put signal.
Selection:
-50 ... +50 %
(in steps of 10 %)
Actual
E1, E2
Display of current start and end potentials
in V.
Scan cycle
Display of current scan cycle mode.
Scan rate
Display of current scan rate in mV/s.
Oven
Display of current oven state.
Range
Display of current measuring range.
Full
scale
Display of current full scale range.
Offset
Display of current offset.
Send current parameters immediately to the
817 Bioscan. Parameters are not stored in
the system file (
*.smt
) as long as the file is
not saved.