6.5 817 Bioscan

IC Net 2.1

73

Method data

Cell potential

Definition of start and end potential for a
user-defined potential sweep, which is ap-
plied to the amperometric detector cell.

E1

Start potential for sweep.
Range:

-2 ... +2 V

.

E2

End potential for sweep.
Range:

-2 ... +2 V

.

Scan cycle

Definition of scan cycle:

full

double

scan

E1 ... E2 ... E1

half

single

scan

E1 ... E2

Selection:

full, half

Scan rate

Definition of scan rate in mV/s.
Selection:

1, 2, 5, 10, 20, 50 mV/s

Oven

Switching on/off column oven and setting
operating temperature.
Range:

15 ... 60 °C

Range

Selection of measuring range:

pA

10 ... 5000 pA

.

nA

0.1 ... 50 nA

.

µA

0.01 ... 5 µA

.

Full

scale

The full scale range (operating range) sets
the desired sensitivity for the analog output.
The possible values of the full scale range
depend on the preset measuring

Range

.

Offset

Offset of the zero point for the analog out-
put signal.
Selection:

-50 ... +50 %

(in steps of 10 %)

Actual

E1, E2

Display of current start and end potentials
in V.

Scan cycle

Display of current scan cycle mode.

Scan rate

Display of current scan rate in mV/s.

Oven

Display of current oven state.

Range

Display of current measuring range.

Full

scale

Display of current full scale range.

Offset

Display of current offset.

Send current parameters immediately to the
817 Bioscan. Parameters are not stored in
the system file (

*.smt

) as long as the file is

not saved.