3 more items to report – Metrohm IC Net 2.1 User Manual

Page 172

background image

7 Methods

IC Net 2.1

164

DATA GLP LOG

Analysing data of channel :ch1

Time constant, ms: 100.00; old 100.01

Iterations: 2; Outliers: 1

Sigma: 0.0

Rejection difference: 1.0

Time error position 5, ratio 2.000, 1 missing points: COR-

RECTED

Corrections 1, Unrecovered errors 0

10/11/1999 16:42:04 Metrohm

Status: Processing

Data saved to disk file

c:\progra~1\metrohm\icnet2~1.0\icnet~1\data\JB101617.CHW




7.6.3

More items to report

This part of the

Report options

window contains a list of rarely

used report items that can be included into the report on the
user's choice:

Acquisition

Parameters used for data acquisition from
the from the

General

,

Measure

and

Filters

tabs of the

Method setup

window.

ACQUISITION PARAMETERS

Channels: 1

Method duration: 25.00min

Run duration: 25.01min

Measurements (method): 15006

Measurements (run): 15006

Freq.divisor: 1

Sampling: 10.00 pts/sec

Start delay: sec

Device:

Program before:

Program after:

Spikes filter: Yes

Median filter: No

slit: 0

Gauss filter: No

slit: 0

Integration

Integration parameters and integration
events.

INTEGRATION DEFAULTS

Channel: ch1

Delay: 2.50 min

Width: 15.00 sec

Broadening: 2.00

Slope: 10.00

Asymmetry: 2.00

MinArea: 0.00

MinHeight: 1.00

Rider ratio: 0.00

No. min

1 5.00 Force horizontal base = 0.00

2 7.50 Cancel horizontal base = 0.00