6.4 761 Compact IC
IC Net 2.1
65
Outputs
The
Outputs
tab of the hardware settings window defines remote
output signals to be set automatically if specific events occur.
Event
Events for automatic remote signal output:
Inject
Switching of the injection valve to the "IN-
JECT" position.
Fill
Switching of the injection valve to the "FILL"
position.
Suppressor step
Switching of the suppressor module to the
next position.
Pump stop (P max/min)
Pump stopped because pressure limits are
exceeded.
Hardware error
Hardware error detected at the 761 Com-
pact IC (high-pressure pump, injection
valve, or suppressor not working correctly).
Alarm leak detector
Leak detector has detected solvent in the
instruments interior.
Remote line
Set remote output lines 1}8.
Selection:
0
(line off, inactive, open)
1
(line on, active, 0 V)
p
(pulse output)
*
(leave line in current status)
PowerOn values
Display of the power on startup values for
remote output lines set on the
Hardware
tab.
Pulse length
Length of a pulse in ms.