11 metrohm suppressor module (msm), 1 protection, Protection – Metrohm 881 Compact IC pro – Anion User Manual
Page 91: Figure 38, Pump tubing connection – changing the filter
![background image](/manuals/586991/91/background.png)
■■■■■■■■■■■■■■■■■■■■■■
5 Operation and maintenance
881 Compact IC pro – Anion
■■■■■■■■
81
1
2
3
Figure 38
Pump tubing connection – Changing the filter
1
Tubing olive
2
Filter 6.2821.130
Packaging contains 10 items.
3
Filter housing
Replacing the filter
1 Unscrewing filter screw
■
Screw the filter screw (38-3) out of the tubing olive (38-1) with
the aid of two 6.2621.000 adjustable wrenches.
2 Replacing the filter
■
Remove the old filter (38-2) with tweezers.
■
Place the new filter (38-2) flat in the tubing olive (38-1) with
tweezers.
3 Mounting filter screw
■
Screw the filter screw (38-3) back into the tubing olive (38-1) and
tighten by hand. Then additionally tighten with two 6.2621.000
adjustable wrenches.
5.11
Metrohm Suppressor Module (MSM)
5.11.1
Protection
To protect the suppressor against foreign particles or bacterial growth, a
pump tubing connection with filter (6.2744.180) (see Figure 25, page 47)
must be mounted between the peristaltic pump (see Figure 23, page 45)
and the inlet capillaries of the suppressor.