3 Installation
881 Compact IC pro – Anion
■■■■■■■■
53
4 Connecting the rinsing the solution outlet capillary
■
Guide the other end of the outlet capillary labeled with waste
rins. into a sufficiently large waste container and fasten it there.
5 Connecting the regeneration solution inlet capillary
■
With a short PEEK pressure screw (6.2744.070), fasten the end of
the inlet capillary labeled with regenerant to the pump tubing
connection of the pump tubing which carries the regeneration
solution.
6 Connecting the regeneration solution outlet capillary
■
Guide the other end of the outlet capillary labeled with waste
reg. into a sufficiently large waste container and fasten it there.
3.17
Connecting the instrument
3.17.1
Connecting the instrument to the PC
Note
The instrument must be switched off when connecting the PC.
1 Connect the USB cable
Connect the PC connection socket of the instrument to a USB con-
nector of the computer via the USB cable (6.2151.020).
3.17.2
Connecting the instrument to mains supply
Warning
The power supply unit must not get wet. Protect it against the direct
effect of liquids.
Mains cable
Which mains cable is supplied depends on the location:
■
6.2122.020 with plug SEV 12 (Switzerland, …)
■
6.2122.040 with plug CEE(7), VII (Germany, …)
■
6.2122.070 with plug NEMA 5-15 (USA, …)