6.1 Problems and their solutions

■■■■■■■■■■■■■■■■■■■■■■

94

■■■■■■■■

881 Compact IC pro – Anion

Problem

Cause

Remedy

Precision problems -
significant scatter-
ing of the measured
values

Injection valve – sample
loop.

Check installation of the sample loop (see
Chapter 3.13.1, page 37)
.

Sample – rinsing volume
too low.

Increase rinsing time (see Chapter 5.8, page
77)
.

Injection valve – defective.

Request Metrohm Service.

Unexpected change
to the retention
times in the chroma-
tograms

Separation column –
diminished separating effi-
ciency.

Regenerate separation column (see Chap-
ter 5.12.4, page 89)
.

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 57)
.

Eluent – Gas bubbles in the
eluent.

Check connections of the eluent degasser (see
Chapter 3.9, page 29)
.

High pressure pump –
defective.

Request Metrohm Service.

Vacuum is not being
built

Eluent Degasser – Connec-
tor Vacuum on the rear of
the instrument not (tightly)
sealed.

Seal the connector Vacuum tightly with a
6.1446.040 threaded stopper.