3 Installation
881 Compact IC pro – Anion
■■■■■■■■
25
3.8
Eluent
3.8.1
Connecting eluent bottle
The eluent is aspirated out of the eluent bottle via the eluent aspiration
tubing (8-1).
The eluent aspiration tubing is connected to the eluent degasser (see
Chapter 3.9, page 29). The tubing must be threaded through a suitable
capillary feed-through of the instrument before the other end can be
equipped.
You will require the parts from the following accessories for equipping the
eluent aspiration tubing:
■
6.1602.160 Eluent bottle attachment GL 45
■
6.2744.210 tubing adapter for aspiration filter
■
6.2821.090 aspiration filter
To equip the eluent aspiration tubing proceed as follows:
Assembling eluent aspiration tubing
1 Guide the free end of the eluent aspiration tubing (8-1) out of the
instrument through a suitable capillary feed-through.
2 Installing eluent bottle attachment 6.1602.160
■
Slide tubing nipple (8-2) and O-ring (8-3) onto the eluent aspira-
tion tubing (8-1).
■
Push eluent aspiration tubing (8-1) through the bottle attachment
(8-4) and screw tight.