3 Installation
881 Compact IC pro – Anion
■■■■■■■■
13
■
Plug a tubing olive (6.2744.034) into one end of the pump tubing
(6.1826.320).
Plug a pump tubing connection (6.2744.180) onto the other end
of the pump tubing.
■
Connect one end of the aspirating capillary (6.1803.020) for the
rinsing solution to the tubing olive on the pump tubing using a
short pressure screw (6.2744.070).
Guide the other end of the aspirating capillary out of the instru-
ment through a capillary feed-through, slide it through a bottle
attachment (6.1602.150) and screw the bottle attachment onto
the bottle (6.1608.020) containing the rinsing solution. Ensure
that the end of the aspirating capillary reaches down to the bot-
tom of the bottle.
■
Place the pump tubing into a tubing cartridge.
Place both tubing cartridges into the peristaltic pump.
6 Connecting the MSM
■
Connect the capillary of the MSM labeled with regenerant to the
pump tubing connection of the pump tubing for the regeneration
solution using a short pressure screw (6.2744.070).
■
Connect the capillary of the MSM labeled with rinsing solution
to the pump tubing connection of the pump tubing for the rinsing
solution using a short pressure screw (6.2744.070).
■
Guide the two capillaries of the MSM labeled with waste reg.
and waste rins. out of the instrument through a capillary feed
through to a waste container and fasten them there.
7 Connecting the instrument
■
Connect the instrument to a computer (see Chapter 3.17.1, page
53) on which the software MagIC Net™ is installed using the
USB cable (6.2151.020).
■
Connect the instrument to the mains supply (see Chapter 3.17.2,
page 53).
8 Initial start-up
■
Switch on the PC and start MagIC Net™.
■
Switch on the instrument.
■
Deaerate the high pressure pump (see Chapter 3.10.2, page
33).
■
Set contact pressure of the peristaltic pump (see "Set flow rate",
page 49).