5 Setting lift positions
815 Robotic Soliprep for LC
■■■■■■■■
51
Work position
1 Move the lift upwards by several mm.
2 Fasten a membrane filter to the Luer adapter of the robotic arm.
3 Under Lift position actuate the arrow key [Arrow down] several
times until the membrane filter is tightly placed on the Luer connector.
4 Click on [Configure rack].
5 Under Lift position select the Work position for External 3.
6 Click on the associated [Assign] button.
7 Click on [Close].
8 Move the lift to the Shift position and remove the membrane filter
from the Luer adapter.
5.6
Lift position for the Polytron
Working height
1 Under Device selection
▶ All devices, click on 815 Robotic
USB... and then select the tab Tower 2.
2 Move to sample position 1. Under Rack position, enter Target posi-
tion 1 and click [Start].
3 Move the lift downward until the Polytron is at a suitable height for
sample dispersion. Click on the arrow button [ Arrow down] and also
(if required) [Arrow up] under lift position in order to set a suitable
position.
4 Click on [Configure rack].
5 Under Lift position, select Work position for tower, if this is not
yet selected.