5.4 Lift positions for stripping needles and filters
■■■■■■■■■■■■■■■■■■■■■■
48
■■■■■■■■
815 Robotic Soliprep for LC
5.4
Lift positions for stripping needles and filters
External position 1
1 Under Device selection
▶ All devices, click on 815 Robotic
USB... and then select the tab Tower 1.
2 Fasten a plastic needle or a membrane filter to the Luer adapter of the
transfer robotic arm.
3 Under Robotic arm position actuate the button [Arrow right] until
the Luer adapter is approximately 2 cm in front of the deflector.
4 Move the lift downward until the thick sleeve of the Luer adapter is
approximately 1 cm underneath the deflector. Click on the arrow but-
ton [Arrow down] and also (if required) [Arrow up] under lift
position in order to set a suitable position.
5 Click on [Configure rack].
6 Under Lift position select Swing position External positions.
7 Click on the associated [Assign] button.
8 Under Robotic arm position select External position 1.
9 Click on the associated [Assign] button.
10 Under Lift position select the Work position for External 1.