5.1 Lift position for sample beakers
■■■■■■■■■■■■■■■■■■■■■■
44
■■■■■■■■
815 Robotic Soliprep for LC
5 Setting lift positions
Various lift positions must be defined before process methods are created
for the Soliprep.
The simplest way to accomplish this is by having the individual positions
established in MagIC Net
™ with the aid of the Manual control. The
established position can then be assigned to a predefined lift position.
The sample rack must be charged with all of the inserts.
Preparing
1 Opening the manual control
In the symbol bar of MagIC Net, click the hand symbol.
2 Under Device selection
▶ All devices, click on 815 Robotic
USB... and then select the tab Tower 1.
5.1
Lift position for sample beakers
Working height
1 Move to sample position 1. Under Rack position, enter Target posi-
tion 1 and click [Start].
2 Fasten a plastic needle to the Luer adapter of the transfer robotic arm.
3 Move the lift downward until the needle is at a suitable height for
aspiration of the dispersed sample. Under Lift position click on the
arrow button [Arrow down] and also (if required) [Arrow up] in
order to set a suitable position.
4 Click on [Configure rack].