5.3 Lift positions for special beakers
■■■■■■■■■■■■■■■■■■■■■■
46
■■■■■■■■
815 Robotic Soliprep for LC
5 Move the lift upward and remove the needle.
Picking up the filter
1 Move to rack position 49. Under Rack position, enter Target posi-
tion 49 and click [Start].
2 Place a membrane filter on the established position in the insert.
3 Move the lift downward until the Luer adapter engages the filter and
it is fixed in place. Under Lift position click on the arrow button
[Arrow down] in order to set a suitable position.
The filter must rest up close against the adapter. The lift may not how-
ever be lowered too far, because otherwise the lift drive could become
overloaded and suffer damage.
4 Note the current lift position (in mm).
This lift height must be entered in the method as absolute lift position.
5 Move the lift upward and remove the filter.
5.3
Lift positions for special beakers
Special beaker 1
1 Move to special beaker position 1. Under Rack position, enter Target
position 103 and click [Start].
2 Place a sample beaker on the established position. Lay the cover with
the hole on the sample beaker.
3 Fasten a membrane filter on the Luer adapter of the transfer robotic
arm.
4 Move the lift downward until the filter rests against the cover of the
sample beaker. Under Lift position click on the arrow button [Arrow
down] in order to set a suitable position.
5 Click on [Configure rack].