14 connecting the polytron – Metrohm 815 Robotic Soliprep for LC User Manual
Page 37

■■■■■■■■■■■■■■■■■■■■■■
2 Installation
815 Robotic Soliprep for LC
■■■■■■■■
31
1
2
2
2
1
1
3
6.2740.020
6.1805.060
6.1805.120
1 Replace the three stoppers on the Polytron robotic arm with
6.2740.020 rinsing nozzles.
2 Connect the three rinsing tubings that are connected to the tower 2
distributor to the rinsing nozzles.
3 The preinstalled tubing of the robotic arm is used to add solvent to the
sample. Connect a 6.1805.120 to it and connect the other end of the
tubing to to the port 1 of the Dosino with a 50 mL dosing unit.
2.14
Connecting the Polytron
®
The Polytron
®
PT 1300 D is comprised of a control module and a disper-
sion drive with a dispersion aggregate for mixing and shredding solid
samples. The control device is directed by a PC software, e.g. tiamo via a
serial RS232 connector.
Consult the Polytron user manual for details about setting up the connec-
tions. The Polytron is connected as follows:
1 Connect the connection cable of the dispersion drive to the front side
of the control module.
2 Check the mains voltage setting on the connection socket on the rear
side of the control module. Switch the device off. Plug in the mains
cable and connect with a socket.
3 Connect the 6.2134.040 RS232 connection cable to the 9-pin con-
nection socket on the rear side of the control module. Plug in the other