beautypg.com

Metrohm 830 IC Interface User Manual

Page 42

background image

4 Appendix

Analog signal interface

Function

2 potential-free analog signal inputs per system.

Connector

Dsub connector 9 pin (female)

1

5

6

9

Input voltage range

±

2.5 V

Input amplification

1, 2, 4, 8, 16, 32, 64 (selectable)

Sampling rate

10, 20, 30, 50, 60 measuring pooints/s

Resolution

24 bit (1 LSP = 0.298

µ

V at amplification 1)

Noise

< 20

µ

Vpp

< 3

µ

Vrms (at 0 V input voltage, amplification 1

and sampling rate 10 Hz)

Zero error

±

2.5 mV

Pin assignment

Pin 1,4,5,8,9: GND (ground, connected to

housing)

Pin 2:

Analog 1+

Pin 3:

Analog 1–

Pin 6:

Analog 2+

Pin 7:

Analog 2–

Safety specifications

Construction/testing

according to EN/IEC 61010-1 / UL 3101-1,

protection class 1, degree of protection IP40

Safety directions

The Instructions for Use include information and

warnings which must be heeded by the user to

assure safe operation of the instrument.

Electromagnetic compatibility (EMC)

Emitted interference

Standards met:

- EN/IEC 61326

- EN 55022

- CISPR 22

Immunity to interference

Standards met:

- EN/IEC 61326

- EN/IEC 61000-4-2

- EN/IEC 61000-4-3

- EN/IEC 61000-4-4

- EN/IEC 61000-4-5

- EN/IEC 61000-4-6

- EN/IEC 61000-4-8

- EN/IEC 61000-4-11

- EN/IEC 61000-4-14

- NAMUR

830 IC Interface / 8.830.1003 Instructions for Use

36