3.2 Settings in the "830 IC Interface" window

Program

On the

Program

subpage, program steps including time, program in-

struction and parameter can be entered.

Erste Spalte

Time at which program instruction is applied.

Entry range:

0.0 ... 999.9 min

If no time is entered, the program instruction is

applied together with the last instruction with

time entry.

Zweite Spalte

Program instruction (see below).
In addition to these predefined instructions,

user-defined remote commands can be en-

tered if activated on the

Events configuration

tab.

Dritte Spalte

Parameter for program instruction (see below).

ENABLED

Enable program start (a disabled program is

not started).

Add new program instruction.

Delete selected program instruction.

Test the time program (error messages are dis-

played if program is wrong).

List of program instructions

The following program instructions can be added to the time program

on the

Program

subpage:

Instruc-
tion

Entry Meaning

Events

0, 1, p, *

Set event output lines 1...7 to the desired
values. For entry of the first value, enter

1

,

0

,

p

or

*

. For entry of the other values, move the

cursor in front of the value to be changed and
enter

1

,

0

,

p

or

*

.

Program END,

RESET

The

END

flag can be used to end a program,

especially if the program time should be longer
than the chromatogram duration. Additional
steps after this flag are not allowed. The

RESET

flag is used to reset the parameters to the
system startup values.

830 IC Interface / 8.830.1003 Instructions for Use