1 event output lines, Event output lines – Metrohm 830 IC Interface User Manual
Page 36
3 Operation
3.2.1
Event output lines
The
830 System [1]
or
830 System [2]
window are available in the SYSTEM
window if the
830 System [1]
or
830 System [2]
have been installed with
the
Install existing devices
option. The
830 System [#]
window with pro-
grams and settings for the seven event output lines is opened by se-
lecting the
Open
menu item with the right mouse button or by double-
clicking the
830 icon
added to the
SYSTEM
window. It consists of the
tabs
Program
and
Events setup
.
Time program
On the
Program
tab of the
830 System [#]
window a user-defined time
program for the event output lines of the 830 IC Interface can be en-
tered. This program is started automatically as defined in the
Start mode
window either at the moment the determination is started (
Start with de-
termination
) or at the moment the sample is injected (
Start with inject
).
The
Program
tab contains the two following subpages::
Program
Main time program with all program steps.
Events configuration
Possibility for creation of user-defined remote
commands.
830 IC Interface / 8.830.1003 Instructions for Use
30