2.4 Connection of external instruments

>CONFIG/RS settings
RS control: on

709 IC Pump:

Switch on external control of the 709 IC Pump using the

[ EXT. ] key.

752 IC Pump Unit, 753 IC Suppressor Module,

754 IC Dialysis Unit:

Switch on external control via remote interface using the

[ Remote ] switch.

817 Bioscan:

This instrument is already pre-configured, no additional set-

tings must be made.

761 Compact IC:

This instrument is already pre-configured, no additional set-

tings must be made.

750 IC Autosampler:

Always switch on the 830 IC Interface first and then the 750

Autosampler.

766 IC Sample Processor:

>RS232 settings
handshake: SWchar

>RS232 settings
RS control: on

Triathlon Autosampler:

This instrument is already pre-configured, no additional set-

tings must be made.

812 IC Valve Unit:

This instrument is already pre-configured, no additional set-

tings must be made.

816 IC Eluent Selector:

Switch on external control via RS232 interface using the

[ EXT ] switch..

828 IC Dual Suppresso:

Switch on external control via remote interface using the

[ Remote ] switch.

837 IC Degasser:

This instrument is already pre-configured, no additional set-

tings must be made.

5 Create new system in «IC Net»

Start the «IC Net» program.

Create a new system file with the selected instruments (de-

tails see IC Net Instructions for Use).


830 IC Interface / 8.830.1003 Instructions for Use