3.2 Settings in the "830 IC Interface" window

3.2

Settings in the "830 IC Interface" window

By clicking the 830 icon in the toolbar with the left mouse button or by

clicking this icon with the right mouse button and selecting the

Open

menu item the

830 IC Interface

window for parameter settings is opened.

It consists of three tabs

Init

,

RS settings

, and

Links

.

Init

The

Init

tab of the

830 IC Interface

window contains data acquisition pa-

rameters for the 830 IC Interface.

Label

Optional label to name the interface with

maximum 8 characters.

ROM version

Instrument program version number .

Measuring per second

Number of data points measured per s.

Entry range:

10, 20, 30, 50, 60 points/s

Range

Range for AD converter .
Example: converts

±

2500 mV to 2

24

bits

Entry range:

2500, 1250, 625, 312.5, 156.25, 78.125,

39.062 mV

Send current parameters to 830 IC Interface.


830 IC Interface / 8.830.1003 Instructions for Use

27