4.1 Technical data

Rs232 interfaces

Connector

Dsub connector 9 pin (male)

1

5

9

6

Baud rate

1'200

115'200 (für Dev. 8 1'200

19'200)

Data bits

7, 8

Stop bits

1, 2

Handshake

Xon/Xoff, none

Parity

none, even, odd

FIFO

0

16 Bytes (für Dev. 8 not available)

Pin assignment

Pin 2:

RxD (Received Data) for Dev. 1

4

Pin 3:

TxD (Transmitted Data) for Dev. 1

4

Pin 5:

GND (Signal Ground) for Dev. 1

4

Pin 7:

TxD (Transmitted Data) for Dev. 5

8

Pin 8:

RxD (Received Data) for Dev. 5

8

Pin 9:

GND (Signal Ground) for Dev. 5

8

Remote interface (Events)

Output lines

7 potential-free relay contacts per system for

controlling external instruments.

RUN

COM


Maximum contact load:

30 V / 1 A DC

Input lines

1 digital input per system for external method

start.

+5V

470k

47k

Ext. Start

RUN

COM

Maximum load for external contact:

approx. 0.1 mA at max. cut-off voltage of approx.

5 V

830 IC Interface / 8.830.1003 Instructions for Use