beautypg.com

5 833 ic liquid handling sample prep unit, 833 ic liquid handling sample prep unit, Fig. 8 – Metrohm 830 IC Interface User Manual

Page 21

background image

2.4 Connection of external instruments

2.4.5

833 IC Liquid Handling Sample Prep Unit

Example

Neutralization of an alkaline sample for anion determination with chemi-

cal suppression.

Instruments

The setup of the instruments corresponds to the Modular IC System 6

(MIC 6) enhanced by the IC Liquid Handling Sample Prep Module

833 (2.833.0030).

2.830.0020 830 IC Interface

2.819.0110 819 IC Detector

2.820.0220 820 IC Separation Center with 2 injectors and

column

heater

2.818.0110 818 IC Pump

2.833.0010 833 IC Liquid Handling Pump Unit

2.833.0020 833 IC Liquid Handling Suppressor Unit

2.833.0030 833 IC Liquid Handling Sample Prep Unit

Interconnection

830

833 Suppressor

833 Sample Prep

833 Pump

820

818

819

PC

6.2134.100

6.2134.090

6.2134.090

6.2134.130

6.2143.210

6.2115.070

6.2125.120

6.2125.090

6.2141.110

6.2143.200

6.2128.180

6.2128.180

6.2128.180

766

B (LOOP) A (PCC)

6.2134.040

Fig. 8:

Connection of 833 (Sample Prep Unit) to MIC 6 for Neutralization

830 IC Interface / 8.830.1003 Instructions for Use

15