6.9 Systematic error diagnostics
■■■■■■■■■■■■■■■■■■■■■■
38
■■■■■■■■
2.896.0020 896 Professional Detector – Amperometry
6.9
Systematic error diagnostics
If the causes of a malfunction cannot be found among the problem
descriptions in the above chapters, then proceed systematically as follows:
Systematic error diagnostics
1 Restarting the instrument and the software
■
Switch off the instrument.
■
Close and restart MagIC Net™.
■
Switch the instrument back on again.
If the problem has not yet been localized, continue with Step 2.
2 Performing instrument test with dummy cell
If the problem has not yet been localized, continue with Step 3.
3 Checking software settings
■
Check the method parameters of the detector and reset them to
values that you know will function.
■
Check the measuring range and reset it to values that you know
will function or select a larger measuring range.
■
Check manual changes to the settings and reset them to values
that you know will function.
■
Check manual settings in the time program and reset them to val-
ues that you know will function.
If the problem has not yet been localized, continue with Step 4.
4 Cleaning the measuring cell
■
Switch off the measuring cell.
■
Remove the measuring cell.
■
Clean the measuring cell (see measuring cell manual).
■
Polish the working electrode (see the leaflet for the working elec-
trode).
■
Reinsert the measuring cell.
If the problem has not yet been localized, continue with Step 5.
5 Replacing the reference electrode
If the problem has not yet been localized, continue with Step 6.