6 Troubleshooting
2.896.0020 896 Professional Detector – Amperometry
■■■■■■■■
31
6 Troubleshooting
6.1
Problems with the hardware
Problem
Cause
Remedy
Leak sensor
responds.
Leaking capillary connec-
tion.
Find any leaking capillary connections and seal
them.
Measuring cell leaking.
Screw apart the measuring cell and then reas-
semble it.
The amperometric
detector is not rec-
ognized in the soft-
ware.
IC system – No connection.
■
Check the cable connection.
■
Switch the IC instrument off and on again
(after 15 seconds).
6.2
Problems with the baseline
Problem
Cause
Remedy
Drift of the baseline.
IC system – Thermal equili-
brium not yet attained.
Condition the system with the heater switched
on.
IC system – Leak in the sys-
tem.
Check capillary connections and seal them.
IC system – Eluent is old
(too much CO
2
).
Apply new eluent.
Pulsing baseline.
High pressure pump – con-
taminated valves.
Clean valves (see Chapter Handling and main-
tenance in the manual for the IC instrument).
High pressure pump –
defective piston seal.
Replace piston seals (see Chapter Handling
and maintenance in the manual for the IC
instrument).
High pressure pump –
Quality of the pump is not
sufficient for the selected
sensitivity.
■
Use pulsation absorber.
■
Use a higher-performance high pressure
pump.
■
Reduce sensitivity.
Measuring cell – Air bubble
in the measuring cell.
■
Deaerate the measuring cell.
■
Degas the eluent continuously.