6 Troubleshooting
2.896.0020 896 Professional Detector – Amperometry
■■■■■■■■
37
Problem
Cause
Remedy
Extreme spread of
the peaks in the
chromatogram.
Splitting (dual
peaks)
IC system – Dead volume
at the ends of the separa-
tion column.
Replace separation column.
IC system – Dead volume
in the IC system.
Check the capillary connectors.
Inhibition of the detection
mechanism by the analyte
(with PAD).
■
Dilute the sample.
■
Allow the waveform to run-in better.
■
Adjust the PAD waveform.
Overload the column.
Dilute the sample.
6.8
Other problems
Problem
Cause
Remedy
High background
current.
IC system – Eluent conta-
minated.
Apply new eluent.
Software – Measurement
potential / pulse settings
incorrect.
Optimize parameters.
Very wide peaks through
substances with delayed
elution.
Wait for the complete elution of these sub-
stances.
Unstable tempera-
ture.
The set temperature is too
low.
Set the temperature to at least 8 °C higher
than the highest ambient temperature to be
anticipated.
Current display/
charge display in the
software is frozen.
Measuring cell – electrodes
are either not connected or
not correctly connected.
Connect the electrode connection cables cor-
rectly (see Chapter 4.7, page 25).
Measuring cell – Small air
bubbles in the measuring
cell.
Deaerate the measuring cell (see Chapter 4.6,
page 24).
Measuring cell – Electrode
connection cable defective.
Perform instrument test with dummy cell (see
Chapter 4.1, page 17).