2 removing / mounting the bottle holder, Removing / mounting the bottle holder – Metrohm 896 Professional Detector – Amperometry User Manual
Page 20
3.2 Mounting base tray and bottle holder (optional)
■■■■■■■■■■■■■■■■■■■■■■
12
■■■■■■■■
2.896.0020 896 Professional Detector – Amperometry
1
3
2
1 Place the base tray in such a way that the openings in the base tray
match exactly the screw threads in the bottom of the instrument.
2 Slide the four washers onto the four cylinder screws, insert the
screws and tighten them with the 3 mm hexagon key.
3 Set the instrument up on the base tray.
Stack other instruments in the required order. Mount the bottle holder
(6.2061.100) onto the topmost instrument on the stack (see "Mounting
the bottle holder", page 13).
3.2.2
Removing / mounting the bottle holder
The bottle holder must be removed in case you want to mount another
instrument on top of the IC instrument. Proceed as follows:
Removing the bottle holder
Before you can remove the bottle holder, the following preconditions
must be met:
■
The instrument is switched off.
■
The bottle holder is cleared.
■
Drainage tubing is disconnected from the drainage tubing connection
of the bottle holder.
To remove the bottle holder, you need a 3 mm hexagon key
(6.2621.100).