6.7 Problems with the chromatogram
■■■■■■■■■■■■■■■■■■■■■■
36
■■■■■■■■
2.896.0020 896 Professional Detector – Amperometry
Problem
Cause
Remedy
No measuring sig-
nal.
IC system – No mains cur-
rent.
Check mains connection and voltage.
Peaks cut off at the
top.
Measuring range too
small.
■
Set a less sensitive measuring range.
■
Reduce peak height, e.g. by means of sam-
ple dilution.
6.7
Problems with the chromatogram
Problem
Cause
Remedy
Peak drift with
sugar analysis.
Carbonate absorption in
the eluent.
Use trap-column Metrosep CO3 Trap 1
(6.1015.300).
Peaks have poor res-
olution.
IC system – Diminished
separating efficiency of the
separation column.
■
Regenerate the separation column (see
Chapter Handling and maintenance in the
manual for the IC instrument).
■
Replace the separation column (see Chap-
ter Handling and maintenance in the man-
ual for the IC instrument).
IC system – Eluent is old.
Apply new eluent.
Ionic strength of the sam-
ple or pH of the sample
deviates greatly from the
eluent.
Dilute the sample or optimize the pH of the
sample.
Absorption of analyte at
the electrodes.
Use a suitable combination of electrodes and
eluent.
Unexpected change
to the retention
times in the chroma-
tograms.
IC system – Diminished
separating efficiency of the
separation column.
■
Regenerate the separation column (see
Chapter Handling and maintenance in the
manual for the IC instrument).
■
Replace the separation column (see Chap-
ter Handling and maintenance in the man-
ual for the IC instrument).
IC system – Eluent is old.
Apply new eluent.
Ionic strength of the sam-
ple or pH of the sample
deviates greatly from the
eluent.
Dilute the sample or optimize the pH of the
sample.