beautypg.com

4 shutting down, 2 amperometric detector, 1 maintenance – Metrohm 896 Professional Detector – Amperometry User Manual

Page 37: 2 preheating capillary maintenance, Shutting down, Amperometric detector, Maintenance, Preheating capillary maintenance

background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation and maintenance

2.896.0020 896 Professional Detector – Amperometry

■■■■■■■■

29

5.1.4

Shutting down

If the instrument is shut down for a longer period of time, the entire IC
system must be rinsed as follows to rid it of salts in order to prevent elu-
ent salts from forming crystals which may cause subsequent damage.

Rinse all capillaries and the Dosino (if present) with methanol/ultrapure
water (1:4),

rinse all pump tubings of the peristaltic pump with ultrapure water.

5.2

Amperometric detector

5.2.1

Maintenance

Warning

When rinsing the detector without column, the pressure must not
exceed 5 MPa.

In order to ensure this, set the maximum pressure of the high pressure
pump to 5 MPa in MagIC Net™.

5.2.2

Preheating capillary maintenance

The preheating capillary can become blocked, e.g. if the IC system has
inadvertently been run dry.

To dissolve this blockage, proceed as follows:

Rinsing the preheating capillary

1 Remove the separation column

Remove the separation column from the IC system and replace with
a coupling (6.2744.040).

2 Adjust settings in MagIC Net™

In MagIC Net™, adjust the following settings:

Maximum pressure of the high pressure pump: 5 MPa

Flow rate: < 0.1 mL/min

3 Rinse the system with the same eluent as before the blockage or

with ultrapure water.

The eluent requires sufficient time to trickle through and dissolve the
crystals.