4 shutting down, 2 amperometric detector, 1 maintenance – Metrohm 896 Professional Detector – Amperometry User Manual
Page 37: 2 preheating capillary maintenance, Shutting down, Amperometric detector, Maintenance, Preheating capillary maintenance
![background image](/manuals/586189/37/background.png)
■■■■■■■■■■■■■■■■■■■■■■
5 Operation and maintenance
2.896.0020 896 Professional Detector – Amperometry
■■■■■■■■
29
5.1.4
Shutting down
If the instrument is shut down for a longer period of time, the entire IC
system must be rinsed as follows to rid it of salts in order to prevent elu-
ent salts from forming crystals which may cause subsequent damage.
■
Rinse all capillaries and the Dosino (if present) with methanol/ultrapure
water (1:4),
■
rinse all pump tubings of the peristaltic pump with ultrapure water.
5.2
Amperometric detector
5.2.1
Maintenance
Warning
When rinsing the detector without column, the pressure must not
exceed 5 MPa.
In order to ensure this, set the maximum pressure of the high pressure
pump to 5 MPa in MagIC Net™.
5.2.2
Preheating capillary maintenance
The preheating capillary can become blocked, e.g. if the IC system has
inadvertently been run dry.
To dissolve this blockage, proceed as follows:
Rinsing the preheating capillary
1 Remove the separation column
Remove the separation column from the IC system and replace with
a coupling (6.2744.040).
2 Adjust settings in MagIC Net™
In MagIC Net™, adjust the following settings:
■
Maximum pressure of the high pressure pump: 5 MPa
■
Flow rate: < 0.1 mL/min
3 Rinse the system with the same eluent as before the blockage or
with ultrapure water.
The eluent requires sufficient time to trickle through and dissolve the
crystals.