Metrohm 896 Professional Detector – Amperometry User Manual
Page 40
6.2 Problems with the baseline
■■■■■■■■■■■■■■■■■■■■■■
32
■■■■■■■■
2.896.0020 896 Professional Detector – Amperometry
Problem
Cause
Remedy
IC system – temperature
fluctuations.
■
Switch on column thermostat or column
oven, respectively.
■
Amperometric detector – Connect preheat-
ing capillary
■
Amperometric detector – Attach and close
front cover (see Chapter 4.8, page 27).
Measuring cell – Working
electrode contaminated.
Clean the working electrode (see the leaflet for
the working electrode).
Measuring cell – Measur-
ing cell leaking.
Check capillary connections on the measuring
cell.
IC system – Eluent conta-
minated.
Apply new eluent.
Smooth baseline (no
noise).
Communications problem
between the amperometric
detector and MagIC Net™.
■
Check the seating of the electrode cable.
■
Check the electrode cable with dummy cell
(see Chapter 4.1, page 17).
■
Switch off the instrument, close and restart
MagIC Net™, switch instrument back on.
All of the data lies outside
of the measuring range.
■
Adjust measuring range.
■
Deaerate the measuring cell (see "Deaerat-
ing the measuring cell", page 24).
Short-circuit bridge
between the electrodes.
■
Examine the working electrode for promi-
nent deposits.
■
Polish the working electrode (see the leaflet
for the working electrode).
■
Replace the working electrode .
■
Clean the measuring cell .
■
Check spacers.
Reference electrode worn
out.
Replace the reference electrode .
The cause is not clear.
Perform systematic error diagnostics (see
Chapter 6.9, page 38).
Unexpectedly high
or low baseline.
Pd reference electrode –
Working conditions not yet
achieved.
Equilibrate until the electrode has adjusted to
the new elution conditions (over night).
DC method – Working con-
ditions not yet achieved.
An excessively high baseline is normal at the
start of the equilibration. Equilibrate until the