Metrohm 756 KF Coulometer User Manual
Page 99

4.2 Remote control commands
756/831 KF Coulometer, Instructions of Use
91
4.2.2.89. Setup.Lock.Keyboard ON, OFF
Setup.Lock.Config ON, OFF
Setup.Lock.Parameter ON, OFF
Setup.Lock.SmplData ON, OFF
Setup.Lock.UserMeth.Recall ON, OFF
Setup.Lock.UserMeth.Store ON, OFF
Setup.Lock.UserMeth.Delete ON, OFF
Setup.Lock.Exchange ON, OFF
Setup.Lock.Display ON, OFF
ON means disable the corresponding function:
.Keyboard: Disable all keys of the Coulometer
.Config:
Disable the
.Parameter: Disable the key
.SmplData:
Disable the
.UserMeth.Recall: Disable "recall" in
.UserMeth.Store: Disable "store" in < USER METH > key
.UserMeth.Delete: Disable "delete" in < USER METH > key
.Exchange:
Disable the
.Display:
Disable the display, i.e. it will not be written to by the device
program of the Coulometer and can be operated from the
computer.
4.2.2.90. Setup.Mode.StartWait ON, OFF
Setup.Mode.FinWait ON, OFF
Holding points in the method sequence. If they are "ON", the sequence stops
until "OFF" is sent. Switching the instrument on sets both nodes to OFF:
.StartWait: Holding point right after starting a method (holding point after
AutoInfo !".T.GC").
.FinWait: Holding point at the end a method (holding point after AutoInfo
!".T.F").
4.2.2.91. Setup.SendMeas.SendStatus ON, OFF
Setup.SendMeas.Interval 0.4...4...16200, MPList
.SendStatus: ON means the automatic transmission of measured values (see
4.2.2.94 and 4.2.2.95) in the inputted interval is active.
.Interval: Time interval (in s) for the automatic transmission of associated
measured values defined under points 4.2.2.95 and 4.2.2.96. The
inputted value is rounded off to a multiple of 0.4. The smallest
possible time interval depends on the number of measured values
which have to be sent, on the baud rate, on the load on the inter-
face and on the type of device connection. With "MPList" the
measured values are sent at the time of their entry into the meas-
ured point list.
The automatic transmission is switched on/off with 'SendStatus'.
4.2.2.92. Setup.SendMeas.Select Assembly,
Titrator
Selection of the unit of which the measured values should be sent (4.2.2.95
or 4.2.2.96).