5.3 Problem with an external printer
756/831 KF Coulometer, Instructions of Use
109
5.3 Problem with an external printer
Problem
Questions for remedial action
No characters can be
received on a connected
printer.
•
Are the instruments switched on and cables plugged
in correctly?
•
Is the printer set to ”on-line”?
•
Are baud rate, data bit and parity the same on both
instruments?
•
Is the handshake set properly?
If everything seems to be OK, try to print a report with
the key sequence
If this report is printed out correctly, check if reports are
defined in key
No data transmission and the
display of the Coulometer
shows an error message.
error 42: Transmission error. Is the printer set to "on-
line"? Is the connection cable properly wired?
The received characters are
garbled.
•
Are the RS settings the same on both devices?
•
Has the correct printer been selected?
•
Data transfer has been interrupted on the hardware
side during the printout of a curve. Re-establish
connections and switch printer off/on.
Wrong line spacing.
The printer does not emulate completely the preset
mode. Usually these problems arise with the IBM mode.
Set the printer to a different mode (e.g. Epson).
Printout of titration curve is
not OK. Other reports are
printed OK.
Handshake is necessary for the printout of curves.
•
Is your cable correctly wired? (The DTR of the printer
has to be connected to the CTS of the Coulometer.)
•
Set "HWs" for the handshake of the Coulometer.
Configure the printer such that its DTR is set
(possibly with DIP switches).