4.2 Remote control commands
756/831 KF Coulometer, Instructions for Use
92
4.2.2.93. Setup.SendMeas.Assembly.CyclNo ON, OFF
Setup.SendMeas.Assembly.I ON, OFF
Setup.SendMeas.Assembly.Meas ON, OFF
Setup.SendMeas.Assembly.Pot ON, OFF
Setup.SendMeas.Assembly.IPulse ON, OFF
Setup.SendMeas.Assembly.Bur.V ON, OFF
Selection of the values from Assembly for the output in the set time interval
(see 4.2.2.92):
.CyclNo: Cycle number of the potential measurement. Together with the
cycle time (4.2.2.77), a time frame can be set up.
The cycle number is set to 0 on switching on the instrument and it
is always incremented as long as the instrument remains
switched on.
.I:
Total charge in mA·s associated to the cycle number, e.g.
"667.48".
.Meas:
Measured value in mV associated to the cycle number, e.g.
"104.2".
.Pot:
Voltage at generator electrode associated to the cycle number.
0 means "undefined", 1 means <14 V, 2 means 14...28 V, 3
means >28 V.
.IPulse:
Current of pulse associated to the cycle number.
1 means 100 mA, 2 means 200 mA, 3 means 400 mA.
.Bur.V:
Dosed volume of connected Dosino in ml, e.g. "5.234".
The unit "assembly" must be preset (see 4.2.2.92).
4.2.2.94. Setup.SendMeas.Titrator.CyclNo ON, OFF
Setup.SendMeas.Titrator.Water ON, OFF
Setup.SendMeas.Titrator.Meas ON, OFF
Setup.SendMeas.Titrator.dWaterdt ON, OFF
Setup.SendMeas.Titrator.I ON, OFF
Setup.SendMeas.Titrator.Pot ON, OFF
Setup.SendMeas.Titrator.IPulse ON, OFF
Selection of the values from the titrator which are sent in the set time interval
(see 4.2.2.91):
.CyclNo: Cycle number. Together with the cycle time (4.2.2.78), a time
frame can be set up. The other data belong to the corresponding
cycle number. The cycle number is set to 0 at the start of a
method and it is incremented until the end of the method.
.Water:
Total water associated to the cycle number in ug, e.g. "62.313"
.Meas:
Measured value in mV at the indicator electrode associated to the
cycle number, e.g. "104.2".
.dWaterdt: Rate or drift associated to the cycle number in ug/min, e.g. "23.0".
.I:
Total charge in mA·s associated to the cycle number, e.g.
"667.48".
.Pot:
Voltage at generator electrode associated to the cycle number.
0 means "undefined", 1 means <14 V, 2 means 14...28 V, 3
means >28 V.
.IPulse:
Current of actual pulse associated to the cycle number.
1 means 100 mA, 2 means 200 mA, 3 means 400 mA.
OV will be sent for "overrange".
The unit "titrator" must be preset (see 4.2.2.92).