Metrohm 756 KF Coulometer User Manual
Page 95

4.2 Remote control commands
756/831 KF Coulometer, Instructions of Use
87
Info.ActualInfo.Titrator.Meas read
only
Info.ActualInfo.Titrator.dWaterdt read
only
Info.ActualInfo.Titrator.I read
only
Info.ActualInfo.Titrator.Pot read
only
Info.ActualInfo.Titrator.IPulse read
only
$Q sends the current values.
.CyclNo: Cycle number of the voltage measurement cycle, e.g. "127". From
the cycle number and the cycle time (see 4.2.2.77), a time frame
can be set up. The cycle number is set to 0 on switching on the
instrument and on every start. It is incremented as long as the in-
strument remains switched on.
.Water:
Total water in ug, e.g. "62.313"
.Meas:
Measured value in mV at the indicator electrode, e.g. "104.2".
.dWaterdt: Rate or drift in ug/min, e.g. "23.0".
.I:
Total charge in mA·s, e.g. "667.48".
.Pot:
Voltage at generator electrode.
0 means "undefined", 1 means <14 V, 2 means 14...28 V, 3
means >28 V.
.IPulse:
Current of actual pulse.
1 means 100 mA, 2 means 200 mA, 3 means 400 mA.
OV will be sent for "overrange".
4.2.2.72. Info.ActualInfo.MeasPt.Index read
only
Info.ActualInfo.MeasPt.X read
only
Info.ActualInfo.MeasPt.Y read
only
Info.ActualInfo.MeasPt.Z1 read
only
Info.ActualInfo.MeasPt.Z2 read
only
Info.ActualInfo.EP.Index read
only
Info.ActualInfo.EP.X read
only
Info.ActualInfo.EP.Y read
only
$Q sends the last entry into the measuring point list (.MeasPt) or the last entry
into the list of EP.
.MeasPt.X Time in s, e.g. "14".
.MeasPt.Y Water in ug, e.g. "27.5".
.MeasPt.Z1 Measured value in mV, e.g. "160.3".
.MeasPt.Z2 Rate in ug/min, e.g. "100.5".
.EP.X
Water in ug, e.g. "26.6".
.EP.Y
Measured value in mV, e.g. "98.6".
4.2.2.73. Info.ActualInfo.Oven.HeatTime read
only
Info.ActualInfo.Oven.SampleTemp read
only
Info.ActualInfo.Oven.LowTemp read
only
Info.ActualInfo.Oven.HighTemp read
only
Info.ActualInfo.Oven.GasFlow read
only
Info.ActualInfo.Oven.UnitFlow read
only
$Q sends the current values from a connected KF Oven. If no Oven is con-
nected, the values are empty.
.HeatTime: Heating time of sample in s.
.SampleTemp: Nominal sample temperature in °C.
.LowTemp: Lowest temperature during the sample heating time in °C.
.HighTemp: Highest temperature during the sample heating time in °C.