3.15 Storing determination results and silo calculations

756/831 KF Coulometer, Instructions for Use

50

3.15 Storing determination results and silo calcula-

tions

3.15.1 Storing determination results

If the sample-specific data of the silo memory should be kept after the determination and

supplemented by results, the following entry is necessary:

In the method under :

Assignment of the determination results to C24 and/or C25:

Assignment of determination results

The determination results are assigned in key .

The display texts of the Coulometer are shown to the

left. The values are the default values.

>silo calculations

C24=

C25=

Silo calculations

Assignment to C24 (RSX, H2O, CXX)

Calculated results (RSX), endpoint (H2O) or variables

CXX can be stored as C24.

Same procedure for C25.

Important:

Ensure that there is still sufficient space for storing the results C24 and C25. (In the report

the number of free bytes is shown.) Result name,

value and unit are stored. The memory requirements can be estimated as follows:

Result with text (8 characters) and unit (5 characters):

32 bytes

def

>formula

>silo calculations

>common variables

>report

>mean