3.4 Configuration, key
756/831 KF Coulometer, Instructions of Use
21
run number 0
operator level: standard
start delay 0 s
result display: bold
dev.label.
beeps 1
display value: OFF
program 5.756.0010
Current run number for result output (0...9999)
The sample number is set to 0 when the instrument is
switched on and incremented on every determination.
Operating mode (standard, expert)
Determines the number of inquiries which are
accessible. Operation in the standard mode contains
only a few inquiries and is recommended for routine
applications.
Inquiries which are accessible in the standard mode
are highlighted in gray in these Instructions for Use.
Start delay (0...999 999 s)
Delay time after start of methods. Abort start delay time
with
Type of result display at the end of the determination
(bold, standard)
bold
: the calculated results are displayed in bold
characters.
standard
: displays the whole information, e.g. results,
water, messages etc.
Individual identification of devices (up to 8 ASCII
characters). Is automatically printed in reports.
Number of beeps (1...3, OFF)
when instrument is ready (conditioning OK), end of
titration and Cond.OK, reception of sample data from
the balance and with sample sizes outside the limiting
values.
Display of measured value (ON, OFF)
Display of U-value during conditioning and titration.
Display of program version. At 831: 5.831.0011 ; at 756:
5.756.0012 .
>RS232 settings COM1
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: HWs
Settings of RS232 interface
see also pages 97ff. Identical for COM2.
Baud rate (300, 600, 1200, 2400, 4800, 9600)
Data bit (7, 8)
Stop bit (1, 2)
Parity (even, odd, none)
Handshake (HWs, SWline, SWchar, none)
see page 97.