3.4 Configuration, key
756/831 KF Coulometer, Instructions of Use
17
3.4 Configuration, key
The key
ment-specific data. The set values apply for all modes.
All entries are only possible in the inactive basic status
of the Coulometer.
Two different operating modes are available: standard
mode and expert mode. Inquiries which appear in the
standard mode are highlighted in gray.
Monitoring functions (only in expert mode):
Monitoring the reagent, validation interval, service in-
terval and printout of diagnostic reports.
Peripheral units (only in expert mode):
Selection of printer, balance, PC keyboard, barcode
reader, stirrer control and selection of the COMs for
manual report output.
Auxiliaries:
e.g. selection of operating mode, setting dialog lan-
guage, date, time.
Settings for RS-COM1 and 2 (only in expert mode):
RS parameters for the interfaces.
Report (only in expert mode):
Configuration of the report.
Common Variable (only in expert mode):
Values of the common variables.
The display texts of the Coulometer are shown to the
left. The values are the default values.
>monitoring
reagent: OFF
Monitoring functions
Monitoring the reagent (ON, OFF)
Monitoring is carried out at the end of the titrations and
when the Coulometer is switched on. If a monitoring
function responds the message "change reagent"
appears. The message vanishes when the reagent is
changed automatically or with
can also be cleared with
all counters are reset to zero.
For generator electrodes with diaphragms the katholyte
normally needs to be changed more frequently than
the anolyte.
CONFIG
configuration
>monitoring
>peripheral units
>auxiliaries
>RS232 settings COM1
>RS232 settings COM2
>report
>common variables