beautypg.com

Metrohm 756 KF Coulometer User Manual

Page 136

background image

7.1 Technical specifications

756/831 KF Coulometer, Instructions for Use

128

Remote Input/Output-

Connection for Oven, Oven Sample Processor, robot.

lines

With optional Remote Box:

Connection for barcode reader and PC keyboard

Dosino connection

For automatic reagent exchange

Ambient temperature

Nom. operation range

5 ... 40 °C

Storage

– 20 ... 60 °C

Transport

– 40 ... 60 °C

Safety specifications

Designed and tested in accordance to IEC publication

1010, safety class I. This manual contains information and

warnings which have to be followed by the user to ensure

safe operation and to retain the apparatus in safe condition.

Mains connection
Voltage

100...240

V

±

10 %

Frequency

50 ... 60 Hz

Power consumption

max. 38 W

Fuse

2 x T1H 250 V (only to be replaced by Metrohm Service

using the same type)

Additional electronic overload protection

Dimensions

At 756

At 831

Width

145 mm

145 mm

Height

194 mm

169 mm

Depth

307 mm

307 mm

Weight, including keypad

approx. 4.5 kg