4.2 Remote control commands
756/831 KF Coulometer, Instructions of Use
89
Assembly.Outputs.ResetLines $G
Setting the I/O output lines.
.AutoEOD: The automatic output of the EOD (End of Determination) at the
end of the determination can be switched off. Thus, for example,
in conjunction with a Coulometer several determinations can be
performed in the same beaker. Before AutoEOD is switched on,
line 3 must be set to "OFF".
.SetLines: With $G, all lines are set.
.SetLines.LX: Set the line LX. "active" means setting of a static signal, "inac-
tive" means resetting of the signal, "pulse" means output of a
pulse of app. 150 ms, "OFF" means the line is not operated, see
also page 131.
Warnings:
•
If you have "AutoEOD" to "ON", an active line 3 is set to "inac-
tive" by the EOD pulse.
•
L6 is the line of the activate pulse. An active line 6 is set to "in-
active" by the activate pulse.
•
L5 is the error line. It is continuously controlled by the Coulom-
eter program and can therefore not be set freely.
Line assignments in Coulometer program:
L0
Ready, inactive state
L1 Conditioning
OK
L2
Titration in progress
L3
EOD (End Of Determination)
L4 ---
L5 Error
L6
Activate pulse + can be set in TIP
L7
Pulses for recorder
L8
Connected remote box
L9,10 ---
L11 Change
reagent
L12
Sample size out of limits
L13
Result out of limits
.ResetLines: Lines are set to the inactive status (= high).
4.2.2.80. Assembly.Stirrer.Status ON, OFF
Switching stirrer ON/OFF.
4.2.2.81. Assembly.Bur.Empty
$G, $S, $H, $C
Assembly.Prep
$G, $S, $H, $C
Starts the function "empty" and "preparation" resp. on the connected Dosino.
4.2.2.82. Assembly.Bur.Rates.Forward.Selected
digital
Assembly.Bur.Rates.Forward.Digital 0...150,
max.
Assembly.Bur.Rates.Reverse.Selected
digital
Assembly.Bur.Rates.Reverse.Digital 0...150,
max.
Expelling and aspirating rate in mL/min. "max." means maximum possible rate
with the Exchange Unit in current use.