beautypg.com

Warning, Before use – Yamato Scientific IM400W Shaking incubator User Manual

Page 9

background image

7

Before Use

Installation and Preparation Before Use

WARNING

Do NOT Use Where Flammable or Explosive Gases or Vapors Are Present.

Do NOT operate this product where there are flammable or explosive gases or vapors since this
product is an arcing device--it produces an arc at power-on--and not explosion-proof. Operation of the
product in such an environment may cause a fire or explosion.

Use Protective Grounding.

Always connect the protective grounding to the ground, to prevent electrical shock or fire due to stray
current. Do NOT connect the grounding wire to gas supply piping. This may cause a fire.

Do NOT connect the grounding wire to the grounding for telephone lines or lightning rods. This may
cause a fire or electrical shock.

„

A grounded 3-prong power outlet is
recommended.

„

If there is no grounding terminal

If a grounding terminal is not provided by the
electric facility, a grounding pole (grounding
resistance

100

, Japanese Industrial Standard

Class 3) must be provided. Since this requires
personnel qualified to do electrical work, consult
the dealer from whom you purchased this product
or the nearest Yamato Scientific sales
representative office.

Grounded power

tl t

Power cord