Before Use

Installation and Preparation Before Use

WARNING

Make fine adjustments for precise level installation.

After determining where to install the product, use the four adjusters to support this product securely.
At that point, also use a level or the like to insure that the product is precisely level. A forward
inclination is especially dangerous since the front door may come open and the shelf inside may slide
out from the vibration. After adjusting the adjusters, tighten all the adjusters in place with the lock nuts.

Do not allow the weight of the product to rest on any caster.

After Installation

To avoid injury, take adequate measures to insure that this product will not fall in the event of an
earthquake or from mechanical shock.

Drainage of Water

z

Water that has condensed inside the incubator is drained through the hose in the rear. Include an
adequate-size drain pan

z

The hose is as shown in Figure 1 when it is delivered to you; however, the layout can be changed as
shown in Figure 2 if it is more suitable for your location.

To change the hose layout

1. First make sure that the power cord is unplugged. Detach the rear cover and louver. To detach

the louver, refer to page 23, “Daily Maintenance and Inspection.”

2. Let the end of the hose stick out from the bottom hole as shown in Figure 2.
3. If the hose seems to be too long, you can make it shorter by cutting it with scissors or the like
4. When laying the hose, be careful not to allow it to be in contact with the V pulley or cooling fan.
5. Be careful not to allow the hose to become twisted.
6. To diminish the intrusion of outside air, make a loop in the hose to make a trap.
7. After the change, reattach the rear cover and louver in their original positions.

Figure 1

Figure 2

Trap

Hose

Trap

Hose