Operation
Placing Flasks
Place flasks properly.
Place each flask so its center is aligned with the middle of the clamp and so it sits straight up.
Do not use a flask that does not fit the clamp.
Attach the clamp properly by inserting it to the
neck of the clamp.
PROPER placement
If a clamp ring is not attached properly or if a flask that does not fit the clamp is used, the flask may
come loose from the clamp while shaking, thus breaking the flask.
Centers must be aligned.
Clamp ring
Must NOT tilt.
Clamp
Not clamped
(Seen from above)
IMPROPER: Clamp ring is
not fit down to its base
position.
IMPROPER: Ring is not catching all the
clamp wires.
IMPROPER: tilted clamp
ring