Safety systems and error codes – Yamato Scientific IM400W Shaking incubator User Manual

Page 23

background image

21

Safety Systems and Error Codes

This incubator features safety systems independent of the controller, in addition to the self-diagnostics
function of the controller. The table below shows the causes when the safety systems go into effect
and the countermeasures to be taken.

If an abnormality such as an operational error or equipment failure occurs, the corresponding error code “ErrX”
(where X is a number) appears on the operation panel display and a buzzer sounds. In this case, check the
error code and immediately stop operation.

Safety System

Display

Causes and Countermeasures

Earth leakage
breaker

None

● Leakage or over-current

Contact your nearest service representative office to check for the
cause of shutdown of the breaker.

Automatic
overheating
protector

None

● Heating of sample

Decrease the sample.

„

Error Codes and Corresponding Countermeasures

Error Code

Display

Description

Countermeasure

Disconnection of
temperature sensor

Replace the sensor.

Short-circuiting of
TRIAC

Replace the TRIAC.

Disconnection of
heater

Replace the heater.

If a problem is detected, the error code appears on the temperature indicator, a buzzer sounds, and then the
freezer and heater circuits are shut off.