Safety Systems and Error Codes
This incubator features safety systems independent of the controller, in addition to the self-diagnostics
function of the controller. The table below shows the causes when the safety systems go into effect
and the countermeasures to be taken.
If an abnormality such as an operational error or equipment failure occurs, the corresponding error code “ErrX”
(where X is a number) appears on the operation panel display and a buzzer sounds. In this case, check the
error code and immediately stop operation.
Safety System
Display
Causes and Countermeasures
Earth leakage
breaker
None
● Leakage or over-current
Contact your nearest service representative office to check for the
cause of shutdown of the breaker.
Automatic
overheating
protector
None
● Heating of sample
Decrease the sample.
Error Codes and Corresponding Countermeasures
Error Code
Display
Description
Countermeasure
Disconnection of
temperature sensor
Replace the sensor.
Short-circuiting of
TRIAC
Replace the TRIAC.
Disconnection of
heater
Replace the heater.
If a problem is detected, the error code appears on the temperature indicator, a buzzer sounds, and then the
freezer and heater circuits are shut off.