Warning, Before use – Yamato Scientific IM400W Shaking incubator User Manual
Page 10

8
Before Use
Installation and Preparation Before Use
WARNING
Observe the Power Rating.
Use a power outlet whose rating meets the electric capacity.
Electric capacity: 100 V AC, 9 A for all IK/IM series products
Do not use a multiple-outlet adapter. It may cause overheating.
Avoid Uneven Surfaces.
Install the product on a level surface. Installing on an uneven surface may cause unanticipated
accidents or failures.
Installation Location
Avoid the following locations:
•
Outdoors
•
Where flammable, corrosive, or
hazardous gases are present
•
Where the ambient temperature is
35°C or higher or 5°C or lower
•
Where rapid changes in temperature
occur
•
Humid or dusty place
•
In direct sunlight
•
Where flooring is not strong enough to
support this product, or on a rack or
the like
•
Where there is vibration
•
Nearby equipment which should be
kept away from vibrations
Front
A clearance of at least 30 cm is needed on
both sides and at the rear (see the figure
on the right)