Maintenance – Yamato Scientific IM400W Shaking incubator User Manual
Page 25

23
Maintenance
Daily Maintenance and Inspection
■
Cleaning the Condenser Slats
Dust adhering to the condenser slats degrades freezing ability and may prevent the temperature from
being controlled at the setpoint. To avoid this, clean the condenser approximately once a month.
Cleaning Procedure
(1) Detach the louver on the front as follows.
Unscrew the round-head M4 screws and detach the louver from the front. The louver hangs on the incubator
casing with two hooks. Lift it up and pull it off.
(2) Clean the condenser slats with a vacuum cleaner or the like.
Be careful not to bend the aluminum slats when cleaning them.
(3) Reattach the louver in its original position.
Louver
Hangs on hooks
Hooks
Condenser slat