beautypg.com

After-sales service and guarantee, Troubleshooting – Yamato Scientific IM400W Shaking incubator User Manual

Page 31

background image

29

After-Sales Service and Guarantee

Troubleshooting

Problem

Causes

Countermeasures

(1) Unusual noise is heard during

shaking

1.

The tray is not sitting on the table

properly, or a flask clamp is not set

properly.

2.

Failure of drawer handle


3. Improper tension of the V belt

Set them properly.

Adjust the hook or replace

the handle.

Adjust the tension.

(2) The incubator itself shakes

during shaking.

1. An adjuster is not touching the floor.
2. The floor or rack is not strong enough to

support the incubator.

3.

The floor or rack is not level.


4. The incubator is not installed on a level

surface.

5.

Samples are too heavy.

Adjust the adjuster height.

Change the installation

location.

Change the installation

location.

Adjust the adjuster heights.

Reduce the weight to

under 6 kg(f).

(3) The table does not shake.

1. The power is not turned on.
2. The run/stop key is not pressed.

3. The door is not shut(if the “door safety”

key is switched on).

4. Loosen V belt
5. Foreign matter has gotten jammed in

the table.

6.

Foreign matter has gotten jammed in

the V pulley.

Turn on the power.

Press the key to start the

shaking.

Close the door.

Adjust the tension.

Remove the foreign

matter.

Remove the foreign

matter.

(4) The internal temperature does

not drop.

1. There is too much frost.
2.

The room temperature is too high.

Defrost.

Lower the room

temperature to no more

than 35°C.

This manual is related to the following products: