beautypg.com

MTS Acumen Electrodynamic Test System Load Frame User Guide User Manual

Page 63

background image

Annually

Monthly

Daily

Calendar Time Using 8 Hours Running
Time Rate per Day

MTS

Check Anti-Rotate Mechanism

MTS

Check For Excessive Contamination

Electrical

MTS

Assure Displacement Transducer Is Clean

MTS

Verify Displacement Transducer Is Aligned

MTS

Confirm Reference Transducer Is Aligned

MTS

Verify Reference Transducer Center
Position

MTS

Check Internal Cables For Abrasion

MTS

Check Internal Cable Connectors

MTS

Check Internal Grounding

MTS

Clean Electronic Devices

System

MTS

Check Frame Mounted Control Emergency
Stop

MTS

Check Remote Emergency Stop (If
Present)

MTS

Check System Detects Velocity Limit When
Enabled

MTS

Check Test Area Enclosure - Open Door
Applies Low Velocity

MTS

Check Frame Mounted Control - High
Power Prohibit Functions

MTS

Verify Unlocked Crosshead Generates
Interlock

MTS

Check System Cable Bundle Connectors

MTS

Check System Cable Bundle For Abrasions

MTS

Check Supplemental Ground Attached

MTS

Verify Tuning Parameters Correct For
Stable Control Modes

MTS Acumen™ Electrodynamic Test System 63